Specifications | Accuracy |
---|---|
Bridge Voltage | Constant voltage DC 1V, 2V, 5V, 7.5V, 10V |
Application bridge resistance | 100Ω ~ 1㏀ |
Rated output | Voltage ±10V (Load resistance more than 200Ω) |
Current 4 ~ 20㎃ (Load resistance less than 300Ω) | |
Zero adjustment ranger | 10%, 10rotations |
Gain (Max 1000) | Switch : 1, 10, 100, 400, 1000 times / Span : 1 ~ 11 times |
S/N rate | 54㏈ |
Frequency response | DC ~ 20 ㎑ (-3㏈) |
Low Pass Filter | 10㎐, 100㎐, 1㎑, 10㎑, WB |
Non-linearity | ±0.01% F.S |
Usage temperature range | 0 ~ 60℃ |
Stability (temp. characteristic) | Sensitivity ±0.03%/℃ |
Humidity | Less than 80% RH (No dew condensation) |
Supply power | AC 220V 50/60㎐ (AC 110V available by internal operation) |
Correction value | 200με, 2000με(1000μStrain = 0.5㎷/V) |